Does different orientation of the Medtronic Hall tilting disc valve in the aortic position effect transvalvuler pressure gradients?

Thumbnail
Göster/Aç
Does different orientation of the Medtronic Hall tilting disc valve in the aortic position effect transvalvuler pressure gradients.pdf (3.409Mb)
Tarih
2000
Yazar
Güden, Metin
Akpınar, Belhhan
Sağbaş, Ertan
Sanisoğlu, İlhan
Arpaz, M.
Bayındır, Osman
Özet
The effect of valve orientation during aortic valve replacement (AVR) with the Medtronic Hall (MH) tilting disc valve on transvalvuler pressure gradients was evaluated. 30 patients (pts) undergoing isolated AVR received a MH which was oriented so that the greater opening of the valve faced the non coronary aortic sinus. There were 12 male and 8 female pts in this group with a mean age of 59 +/- 5.2 years. An echocardiographic examination was performed 6 days after the operation and resting transvalvuler gradients were measured (Group 1). The gradients were compared with the results of 30 pts. who had undergone AVR with the same valve previously but with the orientation of the greater opening towards the left or the right coronary sinus (Group 2). In Group 1 the mean valve size was 22 mm. and the measured mean valvuler gradient was 24/13 mmHg (peak/mean). In Group 2 the mean valve size was 22.1 mm. and the measured mean valvuler gradient was 31/20 mmHg (peak/mean). Student's t test was used for statistical analysis and transvalvular gradients in Group 1 were found to be significantly lower than Group 2 (p>005). Implantation with greater opening of the tilting mechanical valve facing the non-coronary sinus seems to result in lower transvalvuler gradients than other orientations.

Kaynak

49th International Congress of the European Society for Cardiovascular Surgery

Sayfalar

287-291

Bağlantı

https://hdl.handle.net/20.500.12469/259

Koleksiyonlar

  • Araştırma Çıktıları / WOS [1518]

Paylaş


DSpace software copyright © 2002-2015  DuraSpace
İletişim | Geri Bildirim
Theme by 
@mire NV
 

 

Göz at

Tüm Akademik ArşivBölümler & KoleksiyonlarYayın Tarihine GöreYazara GöreBaşlığa GöreKonuya GöreTüre GöreDile GöreDepartmana GöreYayıncıya GöreKHAS Yazarına GöreErişim Türüne GöreBu KoleksiyonYayın Tarihine GöreYazara GöreBaşlığa GöreKonuya GöreTüre GöreDile GöreDepartmana GöreYayıncıya GöreKHAS Yazarına GöreErişim Türüne Göre

Hesabım

GirişKayıt

İstatistikler

Google Analitik İstatistiklerini Görüntüle

DSpace software copyright © 2002-2015  DuraSpace
İletişim | Geri Bildirim
Theme by 
@mire NV