Intraspinal metalloma resulting in late paraparesis

dc.contributor.author Tezer, Mehmet
dc.contributor.author Kuzgun, Ünal
dc.contributor.author Hamzaoğlu, Azmi
dc.contributor.author Öztürk, Çağatay
dc.contributor.author Kabukçuoğlu, Fevziye
dc.contributor.author Şırvancı, Mustafa
dc.date.accessioned 2019-06-27T08:00:52Z
dc.date.available 2019-06-27T08:00:52Z
dc.date.issued 2005
dc.description.abstract The metal-related complications caused by orthopedic implants have long been a concern but these problems have been considered mostly in the field of arthroplasty or internal fixation of fractures. The recent prevalence of spinal instrumentation has evoked a similar concern among spine surgeons. Here we present a case of intraspinal metallosis adjacent to the pedicular hook occurring after treatment of vertebral fracture by posterior spinal instrumentation and fusion and causing paraparesis at the 3rd postoperative year. Metallic granulomas can appear around the pedicular hooks as in the reported case. Crevice and fretting corrosion are results at the junctions of rod-screw rod-hook transverse connector rod and other connector rods in modular spinal implants. Adequate usage of transpedicular screws may inhibit the occurrence of such a complication. For this reason further studies are necessary to increase metallic corrosive resistance to inhibit crevice and fretting corrosion. en_US]
dc.identifier.citationcount 31
dc.identifier.doi 10.1007/s00402-005-0802-x en_US
dc.identifier.endpage 421
dc.identifier.issn 0936-8051 en_US
dc.identifier.issn 1434-3916 en_US
dc.identifier.issn 0936-8051
dc.identifier.issn 1434-3916
dc.identifier.issue 6
dc.identifier.pmid 16034644 en_US
dc.identifier.scopus 2-s2.0-23344445560 en_US
dc.identifier.scopusquality Q1
dc.identifier.startpage 417 en_US
dc.identifier.uri https://hdl.handle.net/20.500.12469/126
dc.identifier.uri https://doi.org/10.1007/s00402-005-0802-x
dc.identifier.volume 125 en_US
dc.identifier.wos WOS:000230706000012 en_US
dc.identifier.wosquality Q2
dc.language.iso en en_US
dc.publisher Springer en_US
dc.relation.journal Archives of Orthopaedic and Trauma Surgery en_US
dc.relation.publicationcategory Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı en_US
dc.rights info:eu-repo/semantics/openAccess en_US
dc.scopus.citedbyCount 37
dc.subject Metallosis en_US
dc.subject Pedicular hook instrumentation en_US
dc.subject Neurological complication en_US
dc.subject Paraparesis en_US
dc.title Intraspinal metalloma resulting in late paraparesis en_US
dc.type Article en_US
dc.wos.citedbyCount 31
dspace.entity.type Publication

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Intraspinal metalloma resulting in late paraparesis.pdf
Size:
352.06 KB
Format:
Adobe Portable Document Format
Description: